Invasive ventilation breathing circuits
Evaqua™ 2 technology within Fisher & Paykel Healthcare’s breathing circuits helps minimize mobile condensate in the expiratory limb by allowing water vapor to diffuse through the tubing wall.
View circuits
Invasive ventilation is delivered directly to a patient’s lower airway via an endotracheal or tracheostomy tube.
This mode of delivery:
• bypasses the natural mechanisms of filtration, humidification and warming, typically provided by the upper airways
• inhibits the primary mechanisms of mechanical clearance such as coughing, sneezing, gagging and particle filtration.
The use of a ventilator is required in invasive respiratory support to enable or support lung function and subsequent gas exchange. In a healthy airway, gas is heated and humidified by the upper airway during inspiration and is fully saturated with water vapor at core body temperature (typically 37 °C, 44 mg/L H2O) when it reaches the distal airways.1
As the mechanisms of a healthy airway are bypassed, delivering heated and humidified gas is mandated and widely practiced during invasive respiratory support of infants.1,2
Delivering gases as close as possible to Optimal Humidity (37 °C, 44 mg/L H2O) supports continued mucociliary transport, as well as encourages open, patent airways for effective gas exchange and ventilation.3
F&P and Evaqua 2 are trademarks of Fisher & Paykel Healthcare Limited. For patent information, see www.fphcare.com/ip